HOGON CT-1PLUS multi-function fully automatic titrator is a new generation titrator that combines color titration and automatic potentiometric titration. The instrument has a modular design and consists of three parts: high-precision measuring tube titration device, controller and detection device. In addition to traditional potentiometric titrations such as pH acid-base titrations, redox titrations, precipitation titrations and complexometric titrations, it can also perform traditional titrations that determine the endpoint by color. The instrument can automatically determine the endpoint and realize fixed endpoint titration, dynamic titration, combined cross titration and manual titration functions. It supports automatic stop detection and manual stop detection, and the core titration components have emergency stop protection functions. Optional automatic color determination module, accurate color determination based on robot vision principle for analysis needs where potentiometric titration cannot be effectively performed
Durable and less than 1ul titration control accuracy, prolong the life of the titrator and provide more accurate results
The test report complies with GLP/GMP regulations, and the USB disk stores the PDF test report for anti-counterfeiting. There is no limit to the number of test methods and test records.
The whole machine has modular design, intelligent control, intelligent calculation, and simple operation to complete complex analysis processes.
Modular design
Consists of high-precision measuring tube + control device + detection device
Special Arrangement
Different formulas can be set according to different testing requirements. Multiple calculation formulas can be set according to detection requirements.
HOGON CT-1PLUS multifunctional fully automatic titrator features
Titration mode Potentiometric titration, color titration
Titration type Acid-base titration, redox titration, color titration, precipitation titration, complexometric titration, non-aqueous titration and pH measurement
End point mode Intelligent determination, equivalence point determination, fixed end point determination, equivalence point + range determination, end point determination, manual titration
Measuring range and display accuracy mV: ±2000mV, resolution: 0.1mV, accuracy ±0.2mV
pH: 0 -14pH, resolution: 0.001 pH, accuracy ±0.003PH
Temperature: -5-125℃, resolution: 0.1℃, accuracy ±0.3℃
Titration control accuracy Burette resolution: 20ml measuring tube (standard): 0.001ml; 10ml measuring tube (optional): 0.0005ml
Titration analysis repeatability: 0.2% (20ml burette)
Titration capacity tolerance: 20ml measuring tube: 0.035ml; 10ml measuring tube: 0.025ml
Buret filling time: 50±10s
HMI 7.0 inch screen displays titration curve and differential curve in real time; can be operated by touch screen or mouse
GLP/GMP regulations Electrode calibration settings, calibration cycle records
USB flash drive storage anti-counterfeiting PDF experimental report
Buret calibration function, calibration record
User group and user permission settings and user operation records
data storage 1G (10W)
Input and output interface RS-232 serial port, USB, network
Working environment Temperature: 5 to 40°C; Humidity: Less than 80% RH (non-condensing)
Power supply: AC 100-240 V, 50/60 Hz; Power: 35W
Dimensions and weight Host: length*width*height=360*250*342 (mm)
Stirring device: length*width*height=320*120*357 (mm)
Weight: 10kg
Optional Constant temperature titration vessel, fully closed titration vessel